Screen-printed anodic stripping voltammetric sensor containing HgO for heavy metal analysis SCIE SCOPUS

Cited 59 time in WEB OF SCIENCE Cited 0 time in Scopus
Title
Screen-printed anodic stripping voltammetric sensor containing HgO for heavy metal analysis
Author(s)
Choi, JY; Seo, K; Cho, SR; Oh, JR; Kahng, SH; Park, J
Alternative Author(s)
조성록; 오재룡; 강성현
Publication Year
2001-09-17
Abstract
A screen-printed disposable sensor for the anodic stripping voltammetric (ASV) analysis of heavy metals was prepared by integrating a HgO-modified working electrode, Ag/AgCl reference electrode, and carbon counter electrode on a polycarbonate strip. The working electrode was screen-printed with phenol resin-based carbon ink containing fine particles of mercuric oxide as a built-in mercury precursor. The mercuric oxide particles exposed on the surface were reduced to fine mercury droplets by in situ or pre-cathodic conditioning in order to allow them to behave as heavy metal collectors in the anodic stripping analysis. The reference and counter electrodes were also screen-printed using commercial silver paste ink modified with finely ground AgCl powder and commercial carbon ink, respectively. The analytical characteristics of the sensor were evaluated using Ph and Cd as probe metals. Detailed preparation procedures and analytical characteristics of the sensor are described. (C) 2001 Elsevier Science B.V. All rights reserved.
ISSN
0003-2670
URI
https://sciwatch.kiost.ac.kr/handle/2020.kiost/5853
DOI
10.1016/S0003-2670(01)01219-3
Bibliographic Citation
ANALYTICA CHIMICA ACTA, v.443, no.2, pp.241 - 247, 2001
Publisher
ELSEVIER SCIENCE BV
Subject
TRACE-METALS; ELECTRODES; LEAD
Keywords
anodic stripping voltammetric analysis; heavy metals; screen-printed disposable sensor