Detecting gold by voltammetric handhold systems KCI

Title
Detecting gold by voltammetric handhold systems
Author(s)
김낙주; 최달웅; 유해수; Kyung Lee; 이장현; 이수영; 김태윤
Alternative Author(s)
유해수
Publication Year
2014-09
Abstract
A voltammetric investigation of Au assay was conducted at a low cost, using Nafion and DNA immobilized on a graphite Pencil working electrode (NDP) with a black lead counter and reference. The following optimal parameters were found: 0.4 V amplitude, 500 Hz frequency, -0.7 V initial potential, and 0.015 V increment potential. These optimal conditions were also applied to sand obtained from the river site. The aforementioned technique is simpler and less costly compared to the common voltammetry and spectrophotometric methods.
ISSN
1225-9098
URI
https://sciwatch.kiost.ac.kr/handle/2020.kiost/2725
DOI
10.12925/jkocs.2014.31.3.472
Bibliographic Citation
오일 및 응용과학 학회지, v.31, no.3, pp.472 - 477, 2014
Publisher
한국응용과학기술학회
Keywords
gold; graphite nafion; voltammetry; alluvial gold